The Model DCA-23 Seawater Dechlorination Analyzer from Electro-Chemical Devices (ECD) is a low maintenance measurement solution that requires no chemicals to monitor seawater chlorine levels accurately in a wide range of industrial process and municipal water treatment applications. The single-purpose analyzer monitors chlorine in seawater from 0.0 to 2.00ppm. It measures the pH, temperature and Total Residual Oxidant (TRO) of seawater and converts them into an equivalent ppm chlorine value, which is displayed to conform with existing conventions. It features a minimum detection limit (MDL) of 0.014 ppm and a limit of quantification (LOQ) of 0.044 ppm, making the DCA-23 Analyzer suitable for monitoring in environments with the most stringent dechlorination regulations The analyzer features three sensors, pH, TRO and temperature. The pH sensor measures from 0-14 pH, the TRO sensor measures from ±1500 mV and the temperature sensor measures from 0-100°C. The DCA-23 with its tri-parameter monitoring, is an ideal seawater chlorine analyzer for a wide range of applications.